優(yōu)質供應PCT17144智能微量殘?zhí)繙y定儀的詳細資料:
PCT17144智能微量殘?zhí)繙y定儀
一、概述
本儀器是按照GB/T 17144《石油產品殘?zhí)繙y定法》、《ASTM D4530》和《ISO 10370:1993》的標準要求設計制造的。具有外形美觀,操作方便等特點,其測試結果與GB/T268(康氏法)測試結果*。
PCT17144智能微量殘?zhí)繙y定儀
二、主要技術參數
1、輸入電源:AC 220V±5% 50Hz
2、加熱功率:950W
3、控溫點:500℃
4、控溫精度:±2℃
5、zui多可做試樣數:12個
6、工作壓力:0.15Mpa
7、氣源壓力:一級壓力zui小2MPa ,二級壓力zui大0.5Mpa
8、氮氣流量:150mL/min 600mL/min (按照GB/T 17144-1997)
9、工作時間:約120min
10、工作環(huán)境溫度:≥25℃
11、工作環(huán)境相對濕度:<85%RH
12、外形尺寸:380×410×570mm(長×寬×高)
PCT17144智能微量殘?zhí)繙y定儀
三、主要特點
儀器外形及結構如上圖所示。與傳統的康拉特遜殘?zhí)繙y定儀相比本儀器具有以下幾項特點:
1、只需少量的試樣即可進行試驗:
殘?zhí)苛浚?) <1 >1-5 >5
試樣量(g) 5.0±0.5 0.5±0.1 0.15±0.05
2、可同時測定12個樣品。
3、采用氮氣保護氣流,確保試驗條件不變能夠獲得良好的重復性和再現性。
4、采用氣管導出法排放試驗尾氣,同康氏法測定儀相比對實驗場所的要求更加自由靈活。
5、采用軸流風機進行冷卻,有助于降低儀器溫度。
6、操作僅需按動“啟動”鍵,一切均由儀器自動進行測試,正常時無需任何操作和判斷,試驗結束后,蜂鳴器提示試驗結束。
7、顯示直觀明了,方便觀察。
PCT17144 inligent micro carbon residue tester
One, overview
The apparatus is in accordance with the GB/T 17144 "petroleum products - Determination of carbon residue," "ASTM D4530" and "ISO 10370:1993" standard requirements of design and manufacturing. Has the advantages of beautiful appearance, convenient operation, the test results and the GB/T268 (Kang's method) consistent test results.
Two, the main technical parameters
1, the input power supply: AC 220V + 5% 50Hz
2, the heating power: 950W
3, temperature 500 DEG C:
4, temperature control precision: - 2 DEG C
5, most can be used as specimen number: 12
6, working pressure: 0.15Mpa
7, air pressure: the pressure minimum level 2MPa, two level of maximum pressure for 0.5Mpa
8, the nitrogen flow rate: 150mL/min 600mL/min (according to GB/T 17144-1997)
9, working time: about 120min
10, the working environment temperature: 25 C
11, the environmental relative humidity: <85%RH
12, size: 380 * 410 * 570mm (length * width * height)
Three, the main features
The shape and structure of the instrument are shown. And Kanglatexun carbon residue tester compared with the traditional instrument has the following features:
1, only a small amount of sample can be tested:
The amount of residual carbon (%) < 1 > 1-5 > 5
Sample size (g) 5 + 0.5 0.5 + 0.1 0.15 + 0.05
2, the simultaneous determination of 12 samples.
3, protected with nitrogen gas flow, ensure the constant test conditions to obtain good repeatability and reproducibility.
4, use tracheal derived method of emission test exhaust, with Kang's method determination instrument compared to the experimental location requirements more free and flexible.
5, adopt a shaft for cooling flow fan, helps to reduce the temperature of the instrument.
In 6, operation only needs to press the "start" button, all are the automatic test instrument, the normal without any operation and judgment, after the test, the buzzer tip end test.
7, display intuitive, convenient for observation.
如果你對優(yōu)質供應PCT17144智能微量殘?zhí)繙y定儀感興趣,想了解更詳細的產品信息,填寫下表直接與廠家聯系: |